Sample Holder Receiving Station – SHR 100 – 400
Together with the OmniVac SH Sample Holder series, the SHR 100 Sample Holder Receiving Station provides easy and versatile access to your samples. The pneumatically driven mechanics establish reliable contact for six electrical connections (e.g. type C thermocouple, heater, 2x Bias or type K thermocouple) as well as for the sapphire sphere equipped cooling finger. The maximum heating or sample current is 15 A. All electrical contacts, heating and cooling can be maintained continuously during sample positioning and rotation, which makes the SHR system unique. The sample transfer station is completely build up of non-magnetic components and can be mounted either with fixed room coordinates or for position alignment on a rotatable HPM XYZ manipulator.
SHR 100 Sample Holder Receiving Station
SHR 100 (OmniVac) | |
Maximal temperature | Depending on used sample holder – 900 K with indirect heating through Cu plate – 1300 K with indirect heating through Mo plate – 1500 K with electron beam or direct heating – Up to 2300 K with electron beam heating and reduced cooling capabilities |
Electrical contacts | 2 for thermocouple (Chromel/Constantan or Chromel/Alumel) – 2 for resistive heater – 2 free pins |
Cooling | Cooling finger with sapphire sphere – LN2 (<100K) or LHe cooling |
Sample holder locking mechanic | pneumatic |
Sample Holder – SH 100
The OmniVac SH 100 Sample Holder series provides a highly flexible mounting system for your samples which meets an outstanding range of desired experimental conditions due to their unique design. The six electrical contacts are equipped with a thermocouple, a resistive heater and two free pins for e.g. biasing or resistance measurements. The whole sample plate can be efficiently cooled down by LN2 (down to 88 K) or LHe.
Heating allows for reaching temperatures up to 2300 K. Different heating methods may be applied as e.g. direct or indirect heating, electron bombardment etc. Many non-standard sample holders are available, e.g. quartz balance, Faraday cup, charge neutralization, optical measurements in transmission etc. Modifications to meet your particular needs can easily be realized.
SH 100 Sample Holder (standard design)
Standard sample size | 1’ |
Max. sample size | 200 mm (diameter) |
Six electrical contacts | 2 for thermocouple, 2 for heating, 2 free pins (e.g. bias, resistance measurements etc.) |
Heating | up to 1300 K (indirect) resp. 2300 K (electron beam) |
Cooling | LN2 (< 100 K) or LHe (< 50 K) |
Adjustment | Internal tilt to adjust sample with measurement plane |
Design variations
Direct heating | SH 130, for e.g. flash-heating of Si/SiO2 |
Electron bombardment heating | SH 120, SH 125 (up to 2300 K) |
Interface to other sample holder systems | e.g. Omicron, SH 170, SH 175, RHK-SPM |
Heatable/coolable sample holder for optical measurements in transmission | SH 150 |
Integrated charge compensation | SH 105 |
Related Products:
- SH 100s – capable of handling Omicron and Leybold system samples